Pipette and Volumetric Glassware Calibration
Request calibration for piston-operated pipettes or one-mark and graduated volumetric apparatus. Check the published volume bands and send the nominal volumes you use.
EGAC CAB 041701: accreditation applies to the activities, ranges, methods and locations in the current approved scope. Confirm the exact work with TEQ before ordering.
Define the calibration you need
For dispensing and solution preparation, list each volume you rely on. The EGAC schedule separates piston-operated apparatus, one-mark volumetric apparatus and graduated volumetric apparatus. The appropriate row depends on the apparatus and the volume requested.
These entries specify the permanent laboratory. Agree the delivery arrangements and any cleaning or contamination requirements with TEQ before sending equipment.
The schedule gives distinct methods and CMCs for these apparatus types. For a multi-channel pipette or an unusual dispenser, include the configuration so TEQ can confirm the method, channels, coverage and reporting.
Published EGAC scope entries for these activities
These published entries describe the selected activities. Confirm the current accreditation status, exact instrument, requested points and service location with TEQ before ordering.
Source: EGAC CAB 041701 · Schedule 041701 B · Revision 01 ·
English technical labels are the controlling source transcription. Arabic labels are an informational reading aid. Numerical values, units, boundary symbols and source anomalies are retained as published.
Calibration CMCs are the scope-specific expanded uncertainty: k=2; approximately 95%. They are not a guaranteed uncertainty for every customer instrument. TEQ confirms the achievable uncertainty for the requested work.
Scroll inside the table to read every column. Source notes appear with the relevant instrument.
| Discipline | Instrument / measurand | Range / property | CMC / uncertainty | Reference method | Location as published | Source page |
|---|---|---|---|---|---|---|
| Volume | Calibration of Volumetric Apparatus (One Mark)Calibration | 0.1 mL to 10 mL | 5.0 μL | SOP TEQ-V1 based on ISO 4787 - 2021 | Permanent | Page 3 |
| Volume | Calibration of Volumetric Apparatus (One Mark)Calibration | > 10 mL to 100 mL | 9.0 μL | SOP TEQ-V1 based on ISO 4787 - 2021 | Permanent | Page 3 |
| Volume | Calibration of Volumetric Apparatus (One Mark)Calibration | > 100 mL to 500 mL | 0.07 mL | SOP TEQ-V1 based on ISO 4787 - 2021 | Permanent | Page 3 |
| Volume | Calibration of Volumetric Apparatus (One Mark)Calibration | > 500 mL to 5000 mL | 0.36 mL | SOP TEQ-V1 based on ISO 4787 - 2021 | Permanent | Page 3 |
| Volume | Calibration of Volumetric Apparatus (Graduated)CalibrationThe values 2.9 mL and 5.8 mL are retained exactly as printed; no decimal correction or unit conversion has been inferred. | 0.1 mL to 10 mL | 9.0 μL | SOP TEQ-V1 based on ISO 4787 - 2021 | Permanent | Page 3 |
| Volume | Calibration of Volumetric Apparatus (Graduated)CalibrationThe values 2.9 mL and 5.8 mL are retained exactly as printed; no decimal correction or unit conversion has been inferred. | > 10 mL to 100 mL | 0.01 mL | SOP TEQ-V1 based on ISO 4787 - 2021 | Permanent | Page 3 |
| Volume | Calibration of Volumetric Apparatus (Graduated)CalibrationThe values 2.9 mL and 5.8 mL are retained exactly as printed; no decimal correction or unit conversion has been inferred. | > 100 mL to 500 mL | 2.9 mL | SOP TEQ-V1 based on ISO 4787 - 2021 | Permanent | Page 3 |
| Volume | Calibration of Volumetric Apparatus (Graduated)CalibrationThe values 2.9 mL and 5.8 mL are retained exactly as printed; no decimal correction or unit conversion has been inferred. | > 500 mL to 5000 mL | 5.8 mL | SOP TEQ-V1 based on ISO 4787 - 2021 | Permanent | Page 3 |
| Volume | Calibration of Piston-operated Volumetric ApparatusCalibration | 0.5 μL to 10 μL | 0.07 μL | SOP TEQ-V2 based on ISO 8655-6-2022 | Permanent | Page 3 |
| Volume | Calibration of Piston-operated Volumetric ApparatusCalibration | > 10 μL to 100 μL | 0.17 μL | SOP TEQ-V2 based on ISO 8655-6-2022 | Permanent | Page 3 |
| Volume | Calibration of Piston-operated Volumetric ApparatusCalibration | > 100 μL to 1000 μL | 0.70 μL | SOP TEQ-V2 based on ISO 8655-6-2022 | Permanent | Page 3 |
| Volume | Calibration of Piston-operated Volumetric ApparatusCalibration | > 1 mL to 10 mL | 5.2 μL | SOP TEQ-V2 based on ISO 8655-6-2022 | Permanent | Page 3 |
| Volume | Calibration of Piston-operated Volumetric ApparatusCalibration | > 10 mL to 200 mL | 18 μL | SOP TEQ-V2 based on ISO 8655-6-2022 | Permanent | Page 3 |
What to include in your quotation request
- List the apparatus type, manufacturer, model, quantity and any asset marks.
- Give the nominal capacity or adjustable range, actual test volumes and units. For glassware, report its class and whether it is marked to contain or deliver if shown.
- For a pipette, provide the channel configuration, tip type and any requested repeatability results. State the tolerance, certificate requirements and preferred date.
Pipette and Volumetric Glassware Calibration: FAQs
Answers on scope, service location and preparing your request.