TEQ EGAC Accreditation and Accreditation Certificate

TEQ is accredited by EGAC to ISO/IEC 17025:2017 (CAB No. 041701). The accreditation status of each requested item is confirmed before work begins.

Accreditation body

Egyptian Accreditation Council (EGAC)

Standard

ISO/IEC 17025:2017

CAB No.

041701

TEQ EGAC Accreditation and Accreditation Certificate

TEQ's published EGAC certificate identifies ISO/IEC 17025:2017 accreditation, CAB No. 041701, and an accompanying detailed scope. The certificate cover lists broad calibration and testing groups; it does not define every instrument's accredited range, method or calibration and measurement capability (CMC). Accreditation applies only to the activities, ranges and locations in the current approved scope. Before requesting accredited calibration, ask TEQ to confirm the instrument, measurement range, uncertainty and service location against that scope. EGAC is an ILAC-MRA signatory; international recognition under ILAC-MRA / Global ACI depends on the applicable recognised scope and current accreditation status.

TEQ's capabilities extend beyond the accredited scope to a wide range of mechanical, physical, and electrical instruments. TEQ calibrates these instruments using reference standards traceable to SI units, qualified personnel, and the same quality-management system. TEQ confirms the accreditation status of each item when reviewing a request, before work begins.

Search the published EGAC scope

Issue 05: 10 July 2025 · Revision 01: 16 July 2026 · Retrieved: 28 October 2026. These are published scope entries; confirm the current status, exact instrument, range and service location with TEQ before ordering.

83 of 83 published range/property entries

Scroll inside the table to view every column and entry. Original English labels, units and numerical values remain available alongside the Arabic reading aid.

EGAC CAB 041701 · Calibration CMCs: expanded uncertainty, k=2, approximately 95%. Testing CMCs: not stated.
DisciplineInstrument / measurandRange / propertyCMC / uncertaintyReference methodLocation as publishedSource page
DimensionsCalibration of CalipersCalibrationUp to 300 mm7.0 μmES 313-1:2018PermanentPage 1
DimensionsCalibration of CalipersCalibration≥ 300 mm to 600 mm8.0 μmES 313-1:2018PermanentPage 1
DimensionsCalibration of External MicrometerCalibrationUp to 150 mm1.7 μmES 163:2015PermanentPage 1
DimensionsCalibration of Dial / Digital GaugeCalibrationUp to 30 mm1.4 μmISO 463:2006PermanentPage 1
Pressure & VacuumCalibration of Vacuum Pressure GaugeCalibration-0.95 bar to 0 bar2.8 mbarDKD-R6-1-2014Permanent & On-SitePage 2
Pressure & VacuumCalibration of Pneumatic Pressure GaugeCalibrationUp to 20 bar2.9 mbarDKD-R6-1-2014Permanent & On-SitePage 2
Pressure & VacuumCalibration of Pneumatic Pressure GaugeCalibration>20 bar to 100 bar14 mbarDKD-R6-1-2014Permanent & On-SitePage 2
Pressure & VacuumCalibration of Hydraulic Pressure GaugeCalibrationUp to 700 bar0.1 barDKD-R6-1-2014Permanent & On-SitePage 2
Pressure & VacuumCalibration of Hydraulic Pressure GaugeCalibration>700 bar to 1000 bar0.26 barDKD-R6-1-2014Permanent & On-SitePage 2
Pressure & VacuumCalibration of Differential Pressure GaugeCalibration-200 Pa to 1000 Pa0.01 hPaDKD-R6-1-2014Permanent & On-SitePage 2
VolumeCalibration of Volumetric Apparatus (One Mark)Calibration0.1 mL to 10 mL5.0 μLSOP TEQ-V1 based on ISO 4787 - 2021PermanentPage 3
VolumeCalibration of Volumetric Apparatus (One Mark)Calibration> 10 mL to 100 mL9.0 μLSOP TEQ-V1 based on ISO 4787 - 2021PermanentPage 3
VolumeCalibration of Volumetric Apparatus (One Mark)Calibration> 100 mL to 500 mL0.07 mLSOP TEQ-V1 based on ISO 4787 - 2021PermanentPage 3
VolumeCalibration of Volumetric Apparatus (One Mark)Calibration> 500 mL to 5000 mL0.36 mLSOP TEQ-V1 based on ISO 4787 - 2021PermanentPage 3
VolumeCalibration of Volumetric Apparatus (Graduated)Calibration0.1 mL to 10 mL9.0 μLSOP TEQ-V1 based on ISO 4787 - 2021PermanentPage 3
VolumeCalibration of Volumetric Apparatus (Graduated)Calibration> 10 mL to 100 mL0.01 mLSOP TEQ-V1 based on ISO 4787 - 2021PermanentPage 3
VolumeCalibration of Volumetric Apparatus (Graduated)Calibration> 100 mL to 500 mL2.9 mLSOP TEQ-V1 based on ISO 4787 - 2021PermanentPage 3
VolumeCalibration of Volumetric Apparatus (Graduated)Calibration> 500 mL to 5000 mL5.8 mLSOP TEQ-V1 based on ISO 4787 - 2021PermanentPage 3
VolumeCalibration of Piston-operated Volumetric ApparatusCalibration0.5 μL to 10 μL0.07 μLSOP TEQ-V2 based on ISO 8655-6-2022PermanentPage 3
VolumeCalibration of Piston-operated Volumetric ApparatusCalibration> 10 μL to 100 μL0.17 μLSOP TEQ-V2 based on ISO 8655-6-2022PermanentPage 3
VolumeCalibration of Piston-operated Volumetric ApparatusCalibration> 100 μL to 1000 μL0.70 μLSOP TEQ-V2 based on ISO 8655-6-2022PermanentPage 3
VolumeCalibration of Piston-operated Volumetric ApparatusCalibration> 1 mL to 10 mL5.2 μLSOP TEQ-V2 based on ISO 8655-6-2022PermanentPage 3
VolumeCalibration of Piston-operated Volumetric ApparatusCalibration> 10 mL to 200 mL18 μLSOP TEQ-V2 based on ISO 8655-6-2022PermanentPage 3
BalanceCalibration of Balances and Weighing InstrumentsCalibrationUp to 22 g10 μgOIML R76-1:2006On-SitePage 4
BalanceCalibration of Balances and Weighing InstrumentsCalibration>22 g to 220 g30 μgOIML R76-1:2006On-SitePage 4
BalanceCalibration of Balances and Weighing InstrumentsCalibration>220 g to 1000 g0.24 mgOIML R76-1:2006On-SitePage 4
BalanceCalibration of Balances and Weighing InstrumentsCalibration>1 kg to 10 kg4.0 mgOIML R76-1:2006On-SitePage 4
BalanceCalibration of Balances and Weighing InstrumentsCalibration>10 kg to 32 kg0.01 gOIML R76-1:2006On-SitePage 4
BalanceCalibration of Balances and Weighing InstrumentsCalibration>32 kg to 100 kg2.0 gOIML R76-1:2006On-SitePage 4
BalanceCalibration of Balances and Weighing InstrumentsCalibration>100 kg to 300 kg10 gOIML R76-1:2006On-SitePage 4
BalanceCalibration of Balances and Weighing InstrumentsCalibration>300 kg to 2000 kg0.10 kgOIML R76-1:2006On-SitePage 4
MassCalibration of MassCalibration1 mg to 20 mg1.3 μgOIML R111:2004/ Amendment : 2025PermanentPage 5
MassCalibration of MassCalibration50 mg to 200 mg2.0 μgOIML R111:2004/ Amendment : 2025PermanentPage 5
MassCalibration of MassCalibration500 mg2.5 μgOIML R111:2004/ Amendment : 2025PermanentPage 5
MassCalibration of MassCalibration1 g3.2 μgOIML R111:2004/ Amendment : 2025PermanentPage 5
MassCalibration of MassCalibration2 g to 5 g5.0 μgOIML R111:2004/ Amendment : 2025PermanentPage 5
MassCalibration of MassCalibration10 g to 20 g8.0 μgOIML R111:2004/ Amendment : 2025PermanentPage 5
MassCalibration of MassCalibration50 g13 μgOIML R111:2004/ Amendment : 2025PermanentPage 5
MassCalibration of MassCalibration100 g19 μgOIML R111:2004/ Amendment : 2025PermanentPage 5
MassCalibration of MassCalibration200 g32 μgOIML R111:2004/ Amendment : 2025PermanentPage 5
MassCalibration of MassCalibration500 g to 1000 g0.13 mgOIML R111:2004/ Amendment : 2025PermanentPage 5
MassCalibration of MassCalibration2 kg1.0 mgOIML R111:2004/ Amendment : 2025PermanentPage 5
MassCalibration of MassCalibration5 kg1.3 mgOIML R111:2004/ Amendment : 2025PermanentPage 5
MassCalibration of MassCalibration10 kg2.5 mgOIML R111:2004/ Amendment : 2025PermanentPage 5
MassCalibration of MassCalibration20 kg9.0 mgOIML R111:2004/ Amendment : 2025PermanentPage 5
TemperatureCalibration of temperature system (Sensors with indicator)Calibration-40 °C to < -25°C0.26 °CSOP TEQ – TC1 based on Euramet cg-08/Ver. 3.1 (02/2020)Permanent & On-SitePage 6
TemperatureCalibration of temperature system (Sensors with indicator)Calibration> -25°C to to 150°C0.06 °CSOP TEQ – TC1 based on Euramet cg-08/Ver. 3.1 (02/2020)Permanent & On-SitePage 6
TemperatureCalibration of temperature system (Sensors with indicator)Calibration> 150°C to 650 °C0.26 °CSOP TEQ – TC1 based on Euramet cg-08/Ver. 3.1 (02/2020)Permanent & On-SitePage 6
TemperatureCalibration of temperature system (Sensors with indicator)Calibration> 650 °C to 900 °C0.61 °CSOP TEQ – TC1 based on Euramet cg-08/Ver. 3.1 (02/2020)Permanent & On-SitePage 6
TemperatureCalibration of temperature system (Sensors with indicator)Calibration> 900 °C to 1200 °C1.6 °CSOP TEQ – TC1 based on Euramet cg-08/Ver. 3.1 (02/2020)Permanent & On-SitePage 6
TemperatureCalibration of Liquid in Glass ThermometerCalibration-25°C to 300 °C0.09 °CSOP TEQ-TG1. In-house method by comparisonPermanent & On-SitePage 6
TemperatureCalibration of Temperature Indicators and Simulator (mv)Calibration-200 °C to 1200 °C0.20 °CSOP TEQ – TS1 based on Euramet cg-11/Ver. 2.0 (03/2011)Permanent & On-SitePage 6
TemperatureCalibration of Temperature Indicators and Simulator (Ohm)Calibration-200 °C to 660 °C0.30 °CSOP TEQ – TS1 based on Euramet cg-11/Ver. 2.0 (03/2011)Permanent & On-SitePage 6
TemperatureCalibration of Block / Well CalibratorCalibration-40 °C to ≤ 150°C0.06 °CSOP TEQ – TBC based on Euramet cg-13/Ver. 4.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of Block / Well CalibratorCalibration>150°C to 650 °C0.20 °CSOP TEQ – TBC based on Euramet cg-13/Ver. 4.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of Block / Well CalibratorCalibration> 650 °C to 900 °C0.60 °CSOP TEQ – TBC based on Euramet cg-13/Ver. 4.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of Block / Well CalibratorCalibration> 900°C to 1200 °C1.5 °CSOP TEQ – TBC based on Euramet cg-13/Ver. 4.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of temperature sensor at working point for closed volumes (freezer, refrigerator)Calibration-80 °C to 10 °C0.20 °CSOP TEQ – TCV based on Euramet cg-20/Ver. 5.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of temperature sensor at working point for closed volumes (incubator, baths, autoclave)Calibration-40°C to 160 °C0.20 °CSOP TEQ – TCV based on Euramet cg-20/Ver. 5.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of a temperature sensor at working point for closed volumes (Oven, muffles, chambers, bathsCalibration25 °C to ≤ 160 °C0.20 °CSOP TEQ – TCV based on Euramet cg-20/Ver. 5.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of a temperature sensor at working point for closed volumes (Oven, muffles, chambers, bathsCalibration> 160°C to 650 °C0.40 °CSOP TEQ – TCV based on Euramet cg-20/Ver. 5.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of a temperature sensor at working point for closed volumes (Oven, muffles, chambers, bathsCalibration> 650°C to 900 °C0.70 °CSOP TEQ – TCV based on Euramet cg-20/Ver. 5.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of a temperature sensor at working point for closed volumes (Oven, muffles, chambers, bathsCalibration> 900°C to 1200 °C1.6 °CSOP TEQ – TCV based on Euramet cg-20/Ver. 5.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of temperature sensor and Humidity at working point for closed volumes (Chamber)Calibration-40°C to 160 °C (20 to 90) %RH0.20 °C 1.8 % RHSOP TEQ – TCV based on Euramet cg-20/Ver. 5.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of Temperature and Humidity Devices (Permanent)Calibration-40 °C to 150°C0.40 °CSOP TEQ – TH based on ES1001-2023PermanentPage 8
TemperatureCalibration of Temperature and Humidity Devices (Permanent)Calibration20% R.H to 90 % R.H1.5 % R.HSOP TEQ – TH based on ES1001-2023PermanentPage 8
TemperatureCalibration of Temperature and Humidity Devices (Permanent)Calibration20 °C to 70°C 20% R.H to 90 % R.H0.40 °C 1.5 % R.HSOP TEQ – TH based on ES1001-2023PermanentPage 8
AccelerometerCalibration of rotational speed devices (rpm) (Mechanical – Photo type – Tachometer)CalibrationUp to 500 rpm0.08 rpmSOP TEQ-R1 based on the Reference Manufacturer ManualPermanent & On-SitePage 9
AccelerometerCalibration of rotational speed devices (rpm) (Mechanical – Photo type – Tachometer)Calibration500 rpm to 1000 rpm0.10 rpmSOP TEQ-R1 based on the Reference Manufacturer ManualPermanent & On-SitePage 9
AccelerometerCalibration of rotational speed devices (rpm) (Mechanical – Photo type – Tachometer)Calibration1000 rpm to 100000 rpm0.82 rpmSOP TEQ-R1 based on the Reference Manufacturer ManualPermanent & On-SitePage 9
TimeCalibration of Stopwatch & TimerCalibrationUp to 24 hr0.19 sec/24 hrSOP TEQ-T1 based on NIST Guide (SP- 960-12)Not stated in the Time sectionPage 10
SpectrophotometerCalibration of UV/Vis SpectrophotometersCalibration200 nm to 900 nm0.31 nmSOP TEQ - SP based on ASTM-E275-8-2022Permanent & On-SitePage 10
SpectrophotometerCalibration of UV/Vis SpectrophotometersCalibration0.1 Abs to 2.5 Abs0.008 AbsSOP TEQ - SP based on ASTM-E275-8-2022Permanent & On-SitePage 10
Conductivity / TDS MeterCalibration of Conductivity DevicesCalibration0 μS/cm to 20000 μS/cm1.0 %SOP TEQ-Cond based on ASTM D1125-14-2023Permanent & On-SitePage 11
Conductivity / TDS MeterTotal Dissolved Solids "TDS" MeterCalibration0 mg/L to 15000 mg/L1.3 %SOP TEQ-Cond based on ASTM D1125-14-2023Permanent & On-SitePage 11
pH MeterCalibration of pH MeterCalibration4.00 7.00 10.000.02SOP TEQ-pH based on ISO 10523-2008.Permanent & On-SitePage 11
ForceCalibration of Universal Testing MachineCalibration60 N to 680 kN0.13 %ISO 7500-1:2018On SitePage 12
Testing — General Products/WorkpiecesDimensions Measurement / DiameterTestingup to 300 mmNot published for this testing entrySOP TEQ – DM based on NPL Guide.PermanentPage 13
Testing — General Products/WorkpiecesDimensions Measurement / LengthTestingup to 5000 mmNot published for this testing entrySOP TEQ – DM based on NPL Guide.PermanentPage 13
Testing — Pressure Safety/Relief ValvesTesting of Operational Hydrilc PressureTestingup to 1200 barNot published for this testing entryISO 4126-1:2013PermanentPage 13
Testing — Water/liquid SampleDetermination of pHTestingup to 14 pHNot published for this testing entrySOP TEQ – WM based on ISO 10523-2008 & ASTM D 1125-14-2023.Not stated for Water/liquid SamplePage 13
Testing — Water/liquid SampleDetermination of ConductivityTestingup to 20 mS/cmNot published for this testing entrySOP TEQ – WM based on ISO 10523-2008 & ASTM D 1125-14-2023.Not stated for Water/liquid SamplePage 13
Testing — Water/liquid SampleDetermination of Total Dissolved Solids (TDS)Testingup to 12000 mg/LNot published for this testing entrySOP TEQ – WM based on ISO 10523-2008 & ASTM D 1125-14-2023.Not stated for Water/liquid SamplePage 13
How to read this scope and its limits
  • Transcribed from the 13-page official EGAC schedule 041701 B, issue 05 (10 July 2025), revision 01 (16 July 2026), retrieved on 28 October 2026. The official record is the source of current status; the listed validity date is not a guarantee against subsequent scope or status changes.
  • This table covers the printed measurands/properties, ranges, CMCs, methods and stated locations on all 13 pages. Range bands are separate rows. It is a transcription aid and does not replace the full schedule or item-specific quotation review. Reference-equipment serial numbers are omitted.
  • Pages 1–12 state expanded uncertainty with k=2 and approximately 95% confidence. The values are CMCs, not a promise of the uncertainty for every customer instrument. No unit conversions, interpolation or recalculation have been applied.
  • Page 13 contains testing entries and does not state CMCs or coverage factors. Empty CMC values for these entries mean not published in that table, not zero uncertainty.
  • Permanent and On-Site qualifiers are retained from the relevant section/row. Where no qualifier is printed, no location is inferred. Working-point temperature-sensor entries do not by themselves establish full chamber mapping or sterilisation validation.
  • The original English labels are the controlling transcription. Arabic labels are informational translations, not an official EGAC Arabic scope. Printed anomalies and boundary symbols are retained in the source fields; consult the official schedule and TEQ before relying on a boundary or ambiguous value.
  • The download endpoint requires a POST request. Open the official EGAC record and use its scope download button; a plain GET link to the endpoint is not a direct PDF download.