On-site calibration brings the agreed measurement work to the customer's premises; laboratory calibration requires the instrument to be sent to the provider's facility. Choose the arrangement that can meet the measurement requirement while fitting equipment access and production needs. Neither location alone proves a better result, an accredited service or shorter downtime. The method, available standards, environment and agreed scope of work must be reviewed for the particular instrument.
Start with removal, access and the measurement requirement
Ask whether the equipment can be removed without disrupting its configuration or creating a difficult recommissioning task. A fixed installation may justify an on-site enquiry; a portable instrument may be convenient to send to a laboratory. These are starting points for discussion, not proof that either service is feasible. Confirm the requested functions, points, acceptance limits and necessary uncertainty before choosing a location.
- On-site may suit an installed or bulky item when suitable reference equipment can reach it and the required conditions can be achieved.
- Laboratory work may suit portable items or measurements needing equipment and environmental control available at the provider's facility.
- Transport may add packing, handling and time away from production; an on-site visit adds access, preparation and technician scheduling requirements.
- A mixed equipment list may need a combined plan, with some instruments evaluated on site and others sent to the laboratory.
- Ask for an item-by-item proposal rather than assigning the whole equipment register to one location without technical review.
Environment and uncertainty must be evaluated
Temperature changes, vibration, air movement, humidity, electrical supply and access can matter differently for different measurements. Describe the actual work area and nearby process activity. Ask the laboratory which conditions it requires, how it will monitor relevant conditions, and what happens if they cannot be maintained. The uncertainty reported for the actual calibration includes relevant contributions from the process and customer's device; the smallest CMC in a scope should not be treated as a guaranteed certificate uncertainty.
A laboratory may offer more control for a particular method, while an on-site measurement may address the instrument in its intended installation. Ask which arrangement answers your question. Calibration of a detached sensor, an indicator using a simulated input and an installed measurement chain are different requests. For temperature equipment, also distinguish an instrument calibration from a chamber study or process validation; those require their own defined objectives and service confirmation.
Check accreditation for the proposed location
An accredited permanent laboratory does not by itself establish that every activity is accredited at a customer's site. EGAC's rules address the use of accreditation references for laboratories holding site calibration accreditation. Obtain the current scope and ask the provider to confirm the specific instrument, range, method and location covered by the proposed accredited work. Service availability at an address and accreditation for that work are separate details to confirm.
If only part of your requested work is covered, require the quotation to identify the status of each part. Do not assume that a discipline such as electrical, mechanical or thermal means every function and range is accredited in both locations. Check any external provider arrangements too. For a full supplier comparison, use How to Choose a Calibration Laboratory in Egypt in the related reading.
Prepare an on-site visit before booking
- Send the site address, equipment locations, models, identifiers, quantities and photographs showing connections and access constraints.
- Agree the points, installed configuration and accessories to be included, and identify any equipment that must be disconnected or made available separately.
- Arrange an appropriate work area, required utilities, access approvals and a site contact; confirm responsibilities for isolation and restoration with your team.
- Share the available shutdown window and allow for setup, stabilisation, measurements, permitted adjustment and completion checks rather than measurement time alone.
- Agree how unsuitable conditions, unavailable items, extra work or a necessary second visit will be handled and charged.
- Confirm reporting, accreditation status and how results before and after any agreed adjustment will be distinguished.
Prepare laboratory work and the return to service
For laboratory work, agree packing and collection or delivery arrangements, include the necessary accessories, and communicate contamination or handling information before shipment. Confirm when the turnaround period starts and whether transport and certificate delivery are included. Provide restrictions on adjustment or repair so unexpected work does not proceed without the approval you require.
After either arrangement, review the certificate against the agreed equipment, points, conditions and reporting needs. For a returned instrument, assess whether transport, reinstallation or a changed configuration calls for checks before use, using the manufacturer's instructions and your quality procedures. The related certificate and traceability guides explain the evidence to review without turning the choice of location into a substitute for that review.
Compare the complete operational cost
Compare like-for-like proposals using the calibration fee, travel or transport, packing, setup, downtime, reporting and any return visit. For example, grouping accessible instruments into one planned shutdown may be practical, while a portable spare can allow an instrument to travel without stopping the process. Ask for confirmed availability and a schedule for your actual site. An on-site enquiry does not guarantee attendance throughout Egypt, a particular response time or that every instrument can be calibrated there.
Official references
- EGAC R4G — Use of accreditation references, including site calibration (clauses 2.2 and 3)
- ISO — ISO/IEC 17025:2017 laboratory competence requirements
- ILAC P14:09/2020 — Background on actual calibration uncertainty and device contributions
- Global ACI — Document register, including TECH-1-009 measurement uncertainty in calibration
Accreditation status for each instrument and range is confirmed by TEQ at quotation stage. The full accredited parameters and ranges are listed in TEQ’s current EGAC scope schedule. TEQ accreditation.
Frequently asked questions
Can on-site calibration be accredited?+
Yes, where the provider's accreditation covers the specific work at the customer's site. Confirm the current scope, instrument, method, range and location with the laboratory before booking. Permanent-laboratory accreditation alone does not establish that coverage.
Is laboratory calibration always more accurate?+
Location alone does not settle suitability. Compare the method, conditions, achievable uncertainty and equipment configuration for the requested measurement. Ask the provider which arrangement can meet your acceptance requirements.
Does on-site calibration eliminate downtime?+
No. Equipment access, isolation, setup, stabilisation and measurement may still interrupt use. Agree the required window and responsibilities in advance, including the plan for items that cannot be completed during the visit.