TEQ · Nasr City, Cairo

Temperature Calibration in Egypt

Request TEQ temperature calibration in Egypt for sensor-and-indicator systems and block/well calibrators. Check published bands, CMCs and service location before booking.

EGAC CAB 041701: accreditation applies to the activities, ranges, methods and locations in the current approved scope. Confirm the exact work with TEQ before ordering.

Define the calibration you need

Define what you need calibrated: the complete sensor-and-indicator system, or a block/well calibrator. Send the probe and readout details, operating points and how the measurement supports your process. The selected published entries below keep each temperature band and CMC separate.

These selected entries state Permanent & On-Site. Confirm whether TEQ should review the complete system in its installed configuration or receive it at the laboratory. Customer-site work depends on the required points, suitable facilities and the agreed method; the quotation confirms availability and conditions.

Calibration of a temperature system is different from chamber temperature mapping, sterilisation validation or process qualification. A single sensor or working-point entry does not establish those activities. The source schedule contains a printed boundary anomaly in one temperature band; it is retained below and must be clarified with TEQ for a request near that boundary.

Published EGAC scope entries for these activities

These published entries describe the selected activities. Confirm the current accreditation status, exact instrument, requested points and service location with TEQ before ordering.

Source: EGAC CAB 041701 · Schedule 041701 B · Revision 01 ·

English technical labels are the controlling source transcription. Arabic labels are an informational reading aid. Numerical values, units, boundary symbols and source anomalies are retained as published.

Calibration CMCs are the scope-specific expanded uncertainty: k=2; approximately 95%. They are not a guaranteed uncertainty for every customer instrument. TEQ confirms the achievable uncertainty for the requested work.

Scroll inside the table to read every column. Source notes appear with the relevant instrument.

Selected EGAC scope entries · 9 published range/property entries
DisciplineInstrument / measurandRange / propertyCMC / uncertaintyReference methodLocation as publishedSource page
TemperatureCalibration of temperature system (Sensors with indicator)CalibrationThe second band repeats "to" in the source. Boundary symbols and numeric values have not been repaired or interpolated.-40 °C to < -25°C0.26 °CSOP TEQ – TC1 based on Euramet cg-08/Ver. 3.1 (02/2020)Permanent & On-SitePage 6
TemperatureCalibration of temperature system (Sensors with indicator)CalibrationThe second band repeats "to" in the source. Boundary symbols and numeric values have not been repaired or interpolated.> -25°C to to 150°C0.06 °CSOP TEQ – TC1 based on Euramet cg-08/Ver. 3.1 (02/2020)Permanent & On-SitePage 6
TemperatureCalibration of temperature system (Sensors with indicator)CalibrationThe second band repeats "to" in the source. Boundary symbols and numeric values have not been repaired or interpolated.> 150°C to 650 °C0.26 °CSOP TEQ – TC1 based on Euramet cg-08/Ver. 3.1 (02/2020)Permanent & On-SitePage 6
TemperatureCalibration of temperature system (Sensors with indicator)CalibrationThe second band repeats "to" in the source. Boundary symbols and numeric values have not been repaired or interpolated.> 650 °C to 900 °C0.61 °CSOP TEQ – TC1 based on Euramet cg-08/Ver. 3.1 (02/2020)Permanent & On-SitePage 6
TemperatureCalibration of temperature system (Sensors with indicator)CalibrationThe second band repeats "to" in the source. Boundary symbols and numeric values have not been repaired or interpolated.> 900 °C to 1200 °C1.6 °CSOP TEQ – TC1 based on Euramet cg-08/Ver. 3.1 (02/2020)Permanent & On-SitePage 6
TemperatureCalibration of Block / Well CalibratorCalibration-40 °C to ≤ 150°C0.06 °CSOP TEQ – TBC based on Euramet cg-13/Ver. 4.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of Block / Well CalibratorCalibration>150°C to 650 °C0.20 °CSOP TEQ – TBC based on Euramet cg-13/Ver. 4.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of Block / Well CalibratorCalibration> 650 °C to 900 °C0.60 °CSOP TEQ – TBC based on Euramet cg-13/Ver. 4.0 (09/2017)Permanent & On-SitePage 7
TemperatureCalibration of Block / Well CalibratorCalibration> 900°C to 1200 °C1.5 °CSOP TEQ – TBC based on Euramet cg-13/Ver. 4.0 (09/2017)Permanent & On-SitePage 7

What to include in your quotation request

  1. Sensor/probe type, manufacturer/model, serial or asset number, readout model and whether the complete measurement chain is included.
  2. Required temperature points, operating range, resolution, tolerance or uncertainty and any conformity requirement.
  3. Probe diameter, immersion depth, connector type and any installation constraints or accessories.
  4. Laboratory or site work, process access, stabilisation time and whether the instrument may be removed from service.
  5. For a chamber enquiry, describe whether you need a sensor check, working-point calibration, mapping or validation. Ask TEQ to confirm the exact activity and accreditation status before ordering.

Temperature Calibration in Egypt: FAQs

Answers on scope, service location and preparing your request.

Should the probe and indicator be calibrated together?
The selected sensor-with-indicator entries concern a temperature system. Tell TEQ whether your request covers that complete system or a separate component; the method, scope entry and reported result must match the agreed work.
Does temperature calibration include chamber mapping?
These are different tasks. State the chamber volume, set points, loading condition and objective if mapping is needed. TEQ must review that activity separately; the selected sensor and block/well entries do not establish accredited mapping coverage.
Can one published CMC be used for every temperature point?
No. The schedule assigns different CMCs to different temperature bands and activities. Read the exact row and its source note. TEQ confirms the uncertainty achievable for your instrument and the requested points.