An out-of-tolerance result means that a stated measurement criterion was not met for the conditions and decision rule used. Start by identifying the instrument, result and rule. Prevent further use for the affected work while the equipment owner evaluates the effect on past measurements and decides what evidence is needed before reuse.
Preserve the evidence and contain use
Keep the original as-found certificate, the instrument identification, service settings, last satisfactory check and any adjustment details. Record the exact point, unit, error and uncertainty. Use the equipment procedure to identify or isolate the equipment and arrange an approved substitute if operations must continue.
Work backwards from the affected measurement
Use the first confirmed unacceptable result and the last comparable satisfactory evidence as starting points. Check whether the instrument was used at the affected points and functions. Review job records, process settings and intermediate checks. An intermediate check may narrow the review window only when it tested a relevant point under comparable conditions.
- List records, lots or decisions that used the instrument during the review period.
- Compare the possible direction and size of the error with the margin available in each decision.
- Record which decisions are supported by independent evidence and which require remeasurement, escalation or further review.
Illustrative example: a balance check
Suppose a balance used near 500 g shows an as-found error of +0.6 g against a two-sided tolerance of ±0.4 g under the agreed rule. Review records taken near that load, including their margin from the acceptance limit and any independent retained checks. These figures are fictional. The appropriate disposition depends on the reported uncertainty, observed use and your own decision rule.
Document the release decision
Record the investigation, results after any authorized adjustment or repair, and the approved return-to-use decision. Link the evidence to the equipment register. Use the observed drift, use conditions, intermediate checks and consequences of error when reviewing the next interval. Record the rationale and the person who approved it.
Official references
Accreditation status for each instrument and range is confirmed by TEQ at quotation stage. The full accredited parameters and ranges are listed in TEQ’s current EGAC scope schedule. TEQ accreditation.
Frequently asked questions
Does a satisfactory as-left result clear past work automatically?+
No. The as-left result describes the instrument after the documented intervention. Use the as-found results and historical evidence to assess work that happened before the intervention.
Does the published CMC give the error on my instrument?+
No. Use the actual reported error and measurement uncertainty on the certificate for the instrument and point you are reviewing.